Alarm Services; Sdn Data Acquisition Service; Recorder Management Services (Recmgr); Documentation Status Services (Docstatusservice) - Philips M3150B Installation And Service Manual

Information center system
Table of Contents

Advertisement

IntelliVue Information Center Application Software
available at the ST/AR output interface. Measurement Services makes this output
available to the Physio Data Server for distribution on request. Measurement
Services also provide the control interface to ST/AR that delivers the specific ST/AR
data requested by an Application.

Alarm Services

Alarm Services provides alarm monitoring for each patient. Each patient has a
separate Alarm Manager that receives its information from the Physio Data Server.
The Alarm Manager determines which events should be announced as alarms or
INOPs and announces them consistent with the bedside-determined priority. For ST/
AR alarms, each Alarm Manager merges ST/AR and bedside monitor alarms and
INOPs, determines the overall priority, and provides the Sound Manager with its
highest priority alarm condition. The Sound Manager causes the tone corresponding
to the most critical condition of all of the patients to be annunciated. For information
on alarm prioritization, see the IntelliVue Information Center User's Guide.
SDN Data Acquisition
SDN Data Acquisition Service is the interface between the Physio Data Server and
Service
the SDN Interface Card in the processing unit. It receives patient data from bedside
and telemetry monitors connected to the SCC through the SDN Interface Card and
transmits them to the Physio Data Server for distribution. It also receives information
from the Physio Data Server and transmits it in SDN message format for display by
bedside monitors.
Recorder Management
Recorder Management Services (RecMgr) provide an interface between the Physio
Services (RecMgr)
Data Server and Documentation Services. For the Philips Recorders, Recorder
Interface Services both transmit data and documentation requests from the
processing unit and bedside monitors to the Recorder and receive status information
from the Recorder for distribution to the processing unit and bedside monitors. Data
communication for both the 2-channel recorder and bedside monitors takes place
through the SDN Interface Card, and for the 4-channel recorder through the Serial B
interface port in the processing unit.
Documentation Status
Documentation Status Services (DocStatus Service) provide documentation control
Services
services to both the Philips Recorder and the LaserJet printer. It receives data from
(DocStatusService)
Recorder Management Services and provides output interfaces to both the Philips
Recorder and the printer through the Windows Print Manager. The data can be either
real-time patient data from bedside or telemetry monitors or stored patient data from
Database Services. Documentation Status Services provide interfaces for data
transmission and to initiate and control printing and recording. A more detailed
description of Documentation Status Services for recording and printing is provided
in the following sections.