Buttons below Monitor/Overview Beds are for assigning beds to Patient Sectors. They
Button
Assign
Note
If all Patient Sectors have been assigned, clicking Change Sector Assignment brings up the following
message.
There are no unassigned sectors left. Please increase the number of sectors if possible, or unassign a sector.
do the following:
Description
Resting Display window opens for making and changing bed assignments to Patient Sectors
The Resting Display window is for assigning patient Beds to Patient Sectors. It has the
following fields.
Available beds lists the Bed labels and clinical Units that have been associated with patient
monitors on this device, but not presently assigned to a Patient Sector.
• For Information Centers that monitor M3/M4/IntelliVue patient monitors, Available
beds are those associated with these monitors in Equipment Setup, but not presently
being monitored by any Information Center.
• For Information Centers that monitor SDN monitors, Available beds are those
associated with SDN monitors in Equipment Setup that share the same SCC as this
device, but not presently being monitored by any Information Center.
• For Information Centers that monitor both M3/M4/IntelliVue and SDN monitors,
Available beds are those associated with both M3/M4/IntelliVue and SDN monitors (that
share the same SCC) in Equipment Setup, but not presently being monitored by any
Information Center.
• For Clients, Available beds are those presently being monitored by an Information Center
but not yet assigned to a Patient Sector on this Client.
• For Patient Link, If there are 12 or less beds configured, only one default CareGroup (#1
Olive) is created, and all Available beds are assigned to this Olive CareGroup. If there are
16 beds configured, two CareGroups (#1 Olive and #2 Teal) are created. In this
configuration, beds assigned to sectors 1-8 are assigned to CareGroup #1 and beds assigned
to sectors 9-16 are assigned to CareGroup #2. The default attributes of the CareGroup(s) are
auto-alarm notification = Red Alarms and Overview prompt sound = enabled.
Software Installation & Configuration
Configuration
6-49