The UPS for the M3154 Database Server is 1000 VA and comes in 3 versions -- 100 VAC
(50-60Hz), 120 VAC (50-60 Hz), and 230 VAC (50-60 Hz). It is shown in Figure 2-26 along
with typical front and rear panels.
l
Test
0
Front Panel
SENSITIVITY
SITE WARNING FAULT
OUTPUT 120-50/60 Hz
1000 Model VA 670W
1500 Model VA 969 W
USB PORT
120V 50-60 Hz
TVSS GND
Rear Panel 100 V
Dip Switches UPSs have rear panel DIP switches that must be set in specific positions for the UPS to
operate properly. Dip switch settings for 120V and 230V models are shown in Figure 2-27.
Note
The switch settings of Figure 2-27 must be made during installation since they are shipped
with all switches on OFF (down).
Test
Switch
On/Off
Switch
Reset
Switch
USB PORT
OVERLOAD
UPS
PROTECTOR
Outputs
INPUT
12A MAX
TVSS GND
Rear Panel 120 V
Figure 2-26 1000 VA UPS for M3154 Database Server
I/O Port
SENSITIVITY
SITE WARNING FAULT
Overload
Protector
Reset
Switch
OUTPUT 120-50/60 Hz
1000 Model VA 670W
1500 Model VA 969 W
UPS
Outputs
INPUT
120V 50-60 Hz
12A MAX
System Components
1000 VA UPS
SENSITIVITY
OUTPUT 220-240-50/60 Hz
1000 Model VA 670W 4.5A
MAX
USB PORT
OVERLOAD
PROTECTOR
INPUT
220V-240 50-60 Hz
7A MAX
TVSS GND
Rear Panel 230 V
2-25