USB Port Test
Required Equipment : Service Dongle
External Port Test
1.) Plug service dongle to the following locations. Refer to
-
-
-
2.) Verify service dongle LED.
Internal Port Test
1.) Remove OPIO Rear Access panel.
2.) Plug service dongle to the following locations.
-
-
3
DRAFT (N
Front USB Port
Service USB Port
Rear USB Access Port
Red/Blue connectors for console without ECG
Black/Blue connectors for ECG equipped console
Without ECG
Figure 4-16 Service Dongle