Replacement of Monitor and LCD Arm Plastic Covers

8-11-1

Manpower

Two person, 15 minutes.
8-11-2

Tools

Phillips screwdriver, Wrench.
8-11-3

Preparations

1.) Lock the Monitor Arm to prevent hazard. Refer to the figure below.
8-11-4

Remove Procedure

1.) Unscrew 2 screws, then remove the Monitor rear panel. Refer to the figure below.
3
DRAFT (N
Figure 8-18 Locking the Monitor Arm
Figure 8-19 Removing the Monitor cable cover
Chapter 8 - Replacement Procedures
17, 2011)
OVEMBER
LOGIQ™ S8
S
M
ERVICE
ANUAL
Locked
2 screws