M-Mode

1.) Press B Mode Button.
2.) Press M Mode Button.
3.) Verify that system scan B/M with the activated probe.
4-3-7-7
B-Flow Mode
NOTE:
Make sure to select probe which supports B-flow
1.) From the touch panel, select Utility -> Imaging -> General.
2.) Change BF/BFC Button to "BF".
3.) Press B-Flow Button.
3
DRAFT (N
Figure 4-25 M-Mode
Figure 4-26 BF Setting
Chapter 4 - Functional Checks
17, 2011)
OVEMBER
LOGIQ™ S8
S
M
ERVICE
ANUAL