Removal Procedure - Trackball; Removal Procedure - Tgc Assy - GE LOGIQ S8 Service Manual

Hide thumbs Also See for LOGIQ S8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
5394227, R
IRECTION
EVISION
8-12-3-4

Removal Procedure - Trackball

1.) Remove connector from back of the PWA (marked
2.) Remove 4 screws around trackball.
3.) Slowly lift up trackball from OP Panel.
NOTE:
LOGIQ S8 Trackball Cable length is 80mm. 210mm cable is for Voluson S8 and not to be used for
LOGIQ S8.
8-12-3-5

Removal Procedure - TGC Assy

1.) Remove TGC Knobs.
2.) Remove connector from back of TGC Assy, marked
3.) Remove 6 screws around TGC Assy.
4.) Slowly lift TGC Assy from OP Panel.
8-32
3
DRAFT (N
Figure 8-32 Trackball
Figure 8-33 TGC Assy
Section 8-12 - Replacement of OPIO and Related Parts
17, 2011)
OVEMBER
below).
below.
LOGIQ™ S8