Removal Procedure - PID, BF192, GFS, CPS

NOTE:
Before working with ACQ box, be sure to remove probes.
1.) Get ACQ Box access.
2.) Remove 9 screws.
3.) Remove EMI Bracket to expose PID, BF192, GFS, and CPS
4.) Draw PID, BF192, GFS, CPS out of the NEST Frame
8-44
3
DRAFT (N
Figure 8-50 ACQ Box
Figure 8-51 Separating Nest Assy