Table 10-14 Type BF Patient Auxiliary Limits
Close Protective earth and close neutral with normal polarity
Close Protective earth and close neutral with reverse polarity
Open Protective earth and close neutral with normal polarity
Open Protective earth and close neutral with reverse polarity
Open neutral and close Protective earth with normal polarity
Open neutral and close Protective earth with reverse polarity
Table 10-15 Isolation between Mains and Applied Limits
Close Protective earth and close neutral with normal polarity
Close Protective earth and close neutral with reverse polarity
NOTE:
*
Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards Table IV
**
Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards, Table
19.5DV.1
*** Measurement limits per IEC 60601-1 Medical Electrical Equipment Safety Standards, Table IV
**** Isolation between Mains and Applied Limits refers to the measurement of leakage current flow
which would flow from mains to patient if the patient were to come into contact with mains voltage.
The following tests are performed at the factory and should be performed at the customer site as well.
-
-
-
-
All measurements should be made with an electrical safety analyzer.
3
DRAFT (N
Conditions
**** Patient Leakage Main to Type BF Applied part
Conditions
Grounding Continuity
Earth Leakage Current.
Enclosure Leakage Current
Patient Leakage Current
Chapter 10 - Care & Maintenance
17, 2011)
OVEMBER
LOGIQ™ S8
S
M
ERVICE
Country
USA
Others
0.1mA*
0.1mA***
0.1mA*
0.1mA***
0.5mA*
0.5mA***
0.5mA*
0.5mA***
0.5mA*
0.5mA***
0.5mA*
0.5mA***
Country
USA
Others
5mA*
5mA***
5mA*
5mA***
ANUAL