Functional Checks

The functional checks take about 60 minutes to perform.
Refer to the LOGIQ™ S8 Basic User Manual whenever necessary.
10-5-2-1

System Checks

Table 10-5
Step
1
2
3
4
5
6
7
8
9
NOTE:
* Some software may be considered standard depending upon system configuration.
3
DRAFT (N
System Functional Checks
Item
Verify basic B Mode (2D) operation.
B Mode
Check the basic system controls that affect this mode of operation.
Verify basic M Mode operation.
M Mode
Check the basic system controls that affect this mode of operation.
Verify basic CFM Mode (Color Flow Mode) operation.
CFM Mode
Check the basic system controls that affect this mode of operation.
Verify basic PD Mode (Power Doppler Mode) operation.
PD Mode
Check the basic system controls that affect this mode of operation.
Verify basic Doppler Mode operation (PW ).
Doppler Modes
Check the basic system controls that affect this mode of operation.
*Applicable Software
Verify the basic operation of all optional modes.
Options
Check the basic system controls that affect each options operation.
Keyboard Test
Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
LCD Monitor
Verify basic LCD Monitor display functions.
Scan a gray scale phantom and use the measurement controls to verify distance and area
Measurements
calculation accuracy.
Refer to the Basic User Manual, for measurement accuracy specifications.