Probe Environmental Requirements
Table 2-5
Temperature
Humidity
Pressure
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
NOTICE
0 TO + 55 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
!! NOTICE:
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5
Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Have two people available to deliver and unpack the LOGIQ™ S8 ultrasound system.
CAUTION
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
!! CAUTION:
result in personal injury and/or damage to the system.
3
DRAFT (N
2D Probe Environmental Requirements
Storage
o
o
0
- 55
C
o
o
32
- 131
F
5 - 85%
non-condensing
700 - 1060hPa
Chapter 2 - Site Preparation
17, 2011)
OVEMBER
Transport
o
o
-40
- 55
C
o
o
-40
- 131
F
5 - 85%
non-condensing
700 - 1060hPa
LOGIQ™ S8
S
M
ERVICE
ANUAL
2-5