7.) Remove 1 screw holding Ground Wire (Green Wire).
8.) Remove 8 screws holding Speaker
8-25-4
Installation
1.) Parts to be installed in reverse order of removal.
8-25-5
Functional Check
Perform Diagsnotics
1.) Enter Common Service Desktop
2.) Select Diagnostics -> Service Diag -> Doppler Audio -> Execute
Make sure audible sound comes out of the speaker
NOTE:
Make sure system volume is not set to zero.
8-96
3
DRAFT (N
Figure 8-128 Removal screws of Speaker