GE H
RAFT
EALTHCARE
D
5394227, R
IRECTION
EVISION
3-6-3
Power Off / Shutdown
After turning off a system, wait at least 10 seconds before turning it on again.
NOTICE
The system may not be able to boot if power is recycled too quickly.
!! NOTICE:
3-6-3-1
Scanner Shutdown
1.) If not already in read mode, freeze the image.
2.) Press the ON/OFF button (see:
3.) Select the SHUTDOWN button. The system performs an automatic full shutdown sequence.
4.) Switch OFF the Circuit at the rear of the system.
NOTE:
The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
ON/OFF button. So the auxiliary equipment need not to be switched ON/OFF separately.
Disconnection of the Main Power Cable is necessary!
WARNING
WARNING
For Example: When repairing the system.
!! WARNING:
5.) After complete power down, disconnect the main power cable from the system or unplug it from the
AC wall outlet socket.
3-24
3
DRAFT (N
Figure
Figure 3-13 System - Exit
Figure 3-14 Circuit and main power at rear of system
Section 3-6 - Completing the Set Up
17, 2011)
OVEMBER
3-11) on the control panel. Following dialog appears.
LOGIQ™ S8
S
ERVICE
Circuit
Main Power
M
ANUAL