Why Do Maintenance; Keeping Records; Maintenance Task Schedule; How Often Should Care & Maintenance Tasks Be Performed - GE LOGIQ S8 Service Manual

Hide thumbs Also See for LOGIQ S8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
5394227, R
IRECTION
EVISION
Section 10-2

Why do Maintenance

10-2-1

Keeping Records

It is good business practice that ultrasound facilities maintain records of quality checks and corrective
maintenance. The Ultrasound Inspection Certificate (see:
documentation that the ultrasound scanner is maintained on a periodic basis.
A copy of the Ultrasound Inspection Certificate should be kept in the same room or near the scanner.
10-2-2
Quality Assurance
In order to gain accreditation from organizations such as the American College of Radiology (USA), it
is the customer's responsibility to have a quality assurance program in place for each scanner.
The program must be directed by a medical physicists, the supervising radiologist/physician or
appropriate designer.
Routine quality control testing must occur regularly. The same tests are performed during each period
so that changes can be monitored over time and effective corrective action can be taken.
Testing results, corrective action and the effects of corrective action must be documented and
maintained on the site.
Your GE service representative can help you with establishing, performing and maintaining records for
a quality assurance program. Please contact us for coverage information and/or price for service.
Section 10-3

Maintenance Task Schedule

10-3-1
How often should care & maintenance tasks be performed?
The Customer Care Schedule (see:
serviced and outlines items requiring special attention.
NOTE:
It is the customer's responsibility to ensure the LOGIQ™ S8 care & maintenance is performed as
scheduled in order to retain its high level of safety, dependability and performance.
Your GE Service Representative has an in-depth knowlegde of your LOGIQ™ S8 ultrasound scanning
system and can best provide competent, efficient service.
Please contact us for coverage information and/or price for service.
The service procedures and recommended intervals shown in the Customer Care Schedule assumes
that you use your LOGIQ™ S8 for an average patient load (10-12 per day) and not use it as a primary
mobile unit which is transported between diagnostic facilities.
NOTE:
If conditions exist which exceed typical usage and patient load, then it is recommended to increase the
maintenance frequencies.
Abbreviations used in the Customer Care Schedule
D = Daily
10-2
3
DRAFT (N
page
W = Weekly
M = Monthly
Section 10-2 - Why do Maintenance
17, 2011)
OVEMBER
page
10-3) specifies how often your LOGIQ™ S8 should be
Table
10-2:
A = Annually
S
ERVICE
10-16) provides the customer with
LOGIQ™ S8
M
ANUAL