About Graphs and Cell results to be displayed
Measurement values of each cell obtained from test results are displayed as graphs in Smart Edge (Circle).
The graph shows the status of edge detection for each cell or whether they are within edge threshold as marks.
It is also possible to shows a list of the status of edge detection or detection data for all cells.
About graph
(1)
Graph type
(2)
Maximum value
(3)
Minimum value
(4)
Angle label
(5)
Result of each cell
170
It shows a graph of radius data when Detect Mode is "Measure Radius". It
shows a graph of diameter data when selecting "Measure Diameter", and a
graph of width data when selecting "Measure Width".
When threshold has been set in Edge Detect Condition, a graph of differential is
shown.
The maximum and minimum values of radius. This is the value of radius
(diameter/width) that is calculated based on the coordinate of edge detected by
each cell before applying various parameters such as Judgement Limits. It may
be different from the maximum measurement or minimum measurement
displayed at the bottom of the screen. (Because the data displayed at the
bottom of the screen is the value after applying the denoising function.)
0 to 360 degree for "Measure Radius" and "Measure Width". 0 to 180 degree for
"Measure Diameter".
The positions and results of edges detected by each cell are displayed from the
start angle in order.
The position at 0 degree and the lower label varies depending on the start angle,
end angle, setting value of "Base Angle" and adjustment angle for
Position/Rotation Adjustment.
It shows detected positions of each cell. It also shows detection results or
maximum and minimum measurement values as marks.
Mark
Color
Description
Yellow green
Min. Measurement
Green
Max. Measurement
×
Red
Edge: Undetected
×
Blue
Excluded edge by
denoising
Light blue
Judgment: OK
Red
Judgement: NG
Cell result list (*)
Display of judgement field
OK/NG
OK/NG
Undetected
Excluded
OK
NG