Panasonic PV500 User Manual page 192

Machine vision system imagechecker
Hide thumbs Also See for PV500:
Table of Contents

Advertisement

Checking result and data of each cell
Press the FUNC key and select "Cell Result" from
the pop-up menu.
Cell number, judgement, measurement values,
coordinate and differential are listed for all the set
cells. The selected cell is displayed in pink on the
screen window.
Tilt the ENTER key up/down to display the data of
the desired cell.
·
In case of the "Measure Width" mode, two sets
of data are displayed for one cell as two
approximate lines are detected.
·
Tilt the SHIFT and ENTER keys to change the
page.
Judgement
Four types of characters are displayed in "Judge." field after the test depending on the edge detection result of
cells or judgement result.
Judgement
Edge Detection
field
1
OK
(Detected)
2
NG
(Detected)
3
Excluded
(Detected)
4
Undetected
(Undetected)
Displaying only the cells of a specific judgement result
It is possible to display only the cell data of a specific result such as the cells judged as NG or the NG and
excluded cells.
Press the FUNC key when displaying the cell result list
and select the type of the cells to be displayed from the
pop-up menu.
The selected result type is checked, and only the cells
of the checked type are displayed in the list. Selecting
again deletes the check marks and restores the
display.
It is possible to select not only one type but also
several types.
For example, when NG cells, undetected cells and
excluded cells are all checked, the cells other than the
cells judged as OK are displayed.
190
Edge coordinate was used
for calculating
measurement value.
×
(Excluded by "Denoising")
×
---
Measurement value is
within the max. and min.
values of Judgement
Limits
(Judgement: OK)
×
(Judgement: NG)
---
---
Mark in
graph
(Light blue)
◆ (Red)
× (Blue)
× (Red)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pv500v2

Table of Contents