Mitsubishi Electric 800 Series Instruction Manual page 417

Hide thumbs Also See for 800 Series:
Table of Contents

Advertisement

• The measurement ends when the timeout time after the maximum slope (Pr.1214) elapsed after the maximum slope is
obtained.
• After the integral term is cleared, PID control is performed with the constant to which the change has been applied (the
constant used before PID gain tuning when a fault occurs).
 Limit cycle method
• In the limit cycle method, the two-position ON/OFF operation is performed three times for output of the manipulated amount
for the real system. From the vibration waveform data of the measured values, the vibration amplitude (Xc) and the
vibration cycle (Tc) are measured. Based on the measured values, each constant is determined.
• In the limit cycle method, less influence of the noise of the measured values is given as compared in the step response
method, and a stable tuning result can be obtained.
Example: Measured value ≤ set point (reverse action)
Set point[%]
Manipulated amount[%]
Limit cycle output upper limit
(Pr.1215-1000)
Limit cycle output lower limit
(Pr.1216-1000)
PID gain tuning start
*1
Details of the two-position ON/OFF operation
PID control
operation
Reverse action
When measured value ≤ set point
Manipulated amount = Upper limit of the output (Pr.1215-
1000)
When measured value > set point
Manipulated amount = Lower limit of the output (Pr.1216-
1000)
Forward action
When measured value ≤ set point
Manipulated amount = Lower limit of the output (Pr.1216-
1000)
When measured value > set point
Manipulated amount = Upper limit of the output (Pr.1215-
1000)
• The manipulated amount is output at the limit cycle output upper limit (Pr.1215 - 1000). (When measured value > set point,
the manipulated amount is once output at the limit cycle lower limit (Pr.1216 - 1000), and then after set point > measured
value is achieved, the manipulated amount is output at the limit cycle output upper limit (Pr.1215 - 1000).)
• The two-position ON/OFF operation is repeated three times. From the waveform data of the values measured for output
of the second and third two-position operation, the vibration amplitude (Xc) and the vibration cycle (Tc) are measured.
• From the vibration amplitude (Xc) and the vibration cycle (Tc), the threshold sensitivity (Ku) and the threshold cycle (Tu)
are calculated.
• Each constant is calculated using a formula depending on the Pr.1218 setting, and PID gain tuning is finished.
• After the integral term is cleared, PID control is performed with the constant to which the change has been applied (the
constant used before PID gain tuning when a fault occurs).
416
5. PARAMETERS
5.11 (A) Application parameters
Measured value[%]
2d
(two-position
output)
During
PID
control
STF
PID
PGT
Initial output of the manipulated amount
Vibration
period Tc
Vibration
amplitude Xc
(1)
(2)
(3)
During PID gain tuning
Two-position ON/OFF operation
Using measured value ≥ set point + hysteresis (Pr.1217)
Manipulated amount = Lower limit of the output (Pr.1216-
1000)
Using measured value ≤ set point - hysteresis (Pr.1217)
Manipulated amount = Upper limit of the output (Pr.1215-
1000)
Using measured value ≥ set point + hysteresis (Pr.1217)
Manipulated amount = Upper limit of the output (Pr.1215-
1000)
Using measured value ≤ set point - hysteresis (Pr.1217)
Manipulated amount = Lower limit of the output (Pr.1216-
1000)
PID gain tuning end
Limit cycle hysteresis
(Pr.1217)
Limit cycle hysteresis
Vibration
amplitude Xc
(Pr.1217)
Time
Two-position
ON/OFF operation
(1) First time
(2) Second time
(3) Third time
Time
During
PID
control
∗1

Advertisement

Table of Contents
loading

Table of Contents