System Performance Verification for HP 4062F
The Performance Verication (PV) program veries the system performance of the HP 4062F.
It is contained in the PV4062.F le. The PV4062.F le is in the /usr/pcs/diag directory.
The PV4062.F program is completely interactive and consists of the following six parts:
4062F DC W/O DVM
4062F DC W/ DVM
4142B
4062F AC
START
VIEW RESULT
After installation, you must perform the 4062F DC W/O DVM, then connection and
calibration of 4062F AC tests. To display or print verication results, use the VIEW RESULT
function. The 4142B and START tests are used for troubleshooting purposes.
Before running PV4062.F, make sure that no system instrument has failed self-test at power
on. For failures, refer to manual of the instrument that failed for instructions on handling
self-test failures.
The HP 4062F needs at least 40 minutes to warm up. If you don't allow the
Note
HP 4062F enough time to warm up, Hewlett-Packard does not guarantee any
performance verication data.
Table 6-3 shows the equipment required to run performance verication.
6-24 System Verification