Figure 4-19 shows the cable connections between the HP 4284A CMU84 and HP 4085F SWM.
Use the following procedures to install the cables:
1. Thread the HP 16048E Test Leads through the left-hand cable entry opening (below the
HP 4284A) on the front panel of the cabinet.
2. Connect the test leads to the UNKNOWN terminals on the front panel of the HP 4284A.
3. Insert the banana terminal of the test leads into the ground terminal (
UNKNOWN terminal.
4. Connect the red and orange leads from the test leads using a BNC-T adapter (HP part
number 1250-0781). Connect the BNC-T adapter to the AUX3 (CMH) terminal of the HP
4085F SWM using a BNC f-to-f adapter (HP part number 1250-0080) and 23 cm BNC
cable (HP part number 8120-1838).
5. Interconnect the gray and black leads from the test leads using a BNC-T adapter. Connect
the BNC-T adapter to the AUX4 (CML) terminal of the HP 4085F SWM using a BNC
f-to-f adapter and 23 cm BNC cable.
The blue lead of the test lead connected to the ground terminal is
Note
necessary for any terminal. Do not connect the terminal of the blue lead to
any terminal.
Figure 4-25. Connecting the HP 4284A CMU84 to the HP 4085F SWM
4-40 System Installation
?
) left side of the
not