shown in Figure 6-44. Con rm that you have prepared everything as described on this
page.
If COM blocks need to be redeclared, START program runs automatically to redeclare
them. If so, con guration page is displayed temporarily, then preparation page is displayed.
d
Read the above message, then press "NEXT" to continue.
aaaaaaaaaaaaaaa
c
2. Install the pulse test module (HP 16354F) on the SWM.
6-42 System Verification
softkey from PV4062.F main menu. Preparation page is displayed as
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
HP 4062F Performance Verification Program for AC function
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
1. Confirm the Pin Boards for 4062F are installed at least at Pin4, 8, 12, 16,
20, 24, 28, 32, 36, 40, 44, and 48 in the SWM.
2. Connect the Probe to HP54121A INPUT 3 as follows:
[HP 54121A INPUT3] <-> [20dB ATT] <-> [SMA-BNC Adapter] <-> [BNC Cable]
3. Remove 50 Ohm terminator from PG1 TRIG OUT cable, then connect the cable
to HP 54121A TRIG terminal.
4. Set the HP-IB address of HP 54120B to 2707 (To change address, select
"HP-IB ADDRESS" softkey.
5. Mount the HP 16354F atop the SWM.
aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa aaaaaaaaaaaaaaa
NEXT
Figure 6-44. Preparation Page for 4062F AC