HP-IB Interface Cards

The HP 4062UX/F requires an HP 98624A Interface Card in addition to the Series 300
built-in HP-IB interface. Each HP-IB interface is reserved for a speci c system function, as
listed in Table 2-1.
Interface
Computer's Built-in HP-IB
HP 98624A
HP 98624A
HP 98624A HP-IB Interface Cards must be con gured as shown Table 2-2.
If you con gure two HP-IB interface cards for the system instruments and the wafer prober,
set the
of one interface card to 27 and the other to 25. If you con gure one HP-IB
select code
interface card for the system instruments, set the select code of the interface card to 27.
Figure 2-1 shows the switch settings of the HP-IB interface card.
2-4 System Controller Installation
Table 2-1. HP-IB Interface Settings
Select Code
7
For communicating with computer peripherals or
controlling other instruments
27
For controlling the system instruments; HP 4084B, HP
4142B, HP 4280A, HP 4284A, and HP 8110As
25
For controlling the HP-IB controlled wafer probers on the
HP 4062UX/F system
Table 2-2. HP-IB Interface Card Switch Settings
Factory Settings
Non-system Controller
Interrupt Level: 3
Select Code: 8
Primary Address: 0
Description
. . .
be changed
System Controller
Interrupt Level: 3
Select Code: 27 or 25
Primary Address: 21