Connections Between Output Terminals Of The Hp 8110A And Hf Ports - HP 4062UX Installation Manual

Semiconductor process control system semiconductor parametric test system
Table of Contents

Advertisement

Figure 4-27 shows the cable connections between the HP 8110A PGs and HP 4085F SWM.
The PG cables are wrapped in a zipper tube and xed on the cabinet frame with cable ties at
the factory. Use the following procedure to connect the PG cables:
1. Attach the cable adapter (HP part number 04085-61066) to the HP 4085F SWM.
2. On rear panel of HP 8110A, connect BNC connector of each PG cable to corresponding
OUTPUT 1 or OUTPUT 2 terminal. Each PG cable has a tag that indicates the
OUTPUT terminal of HP 8110A to which you should connect the cable. For example, if
cable has tag \PG31", connect it to OUTPUT 1 terminal of the HP 8110A (PG3).
3. Pass the PG cables through the opening of the rear panel.
4. On the cable adapter that is attached to HP 4085F SWM, connect BNC connector of each
PG cable to corresponding HF port. Hewlett-Packard recommends that when you install
the HP 4062F system, you connect the PG cables to the HF ports as shown in Table 4-13.

Connections Between Output Terminals of the HP 8110A and HF Ports

HP 8110A Name
PG1
PG2
PG3
PG4
PG5
Figure 4-27. Connecting the HP 8110A PGs to the HP 4085F SWM
4-42 System Installation
Table 4-13.
Output Terminal
OUTPUT 1
OUTPUT 2
OUTPUT 1
OUTPUT 2
OUTPUT 1
OUTPUT 2
OUTPUT 1
OUTPUT 2
OUTPUT 1
OUTPUT 2
PG Cable
HF Port
PG11
PG12
PG21
PG22
PG31
PG32
PG41
PG42
PG51
PG52
HF10
HF1
HF2
HF3
HF4
HF5
HF6
HF7
HF8
HF9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 4062f

Table of Contents