4062F Ac" Test; Required Equipment - HP 4062UX Installation Manual

Semiconductor process control system semiconductor parametric test system
Table of Contents

Advertisement

\4062F AC" Test

The 4062F AC test provides the following ve subtests for verifying the SWM and PG
operation.
Connection Test
Calibration
HP 8110A Self-test
Switch Test
Pulse Test
To verify HP 4062F after installation, you must execute connection test and calibration. So,
this section provides instructions for performing the connection test and calibration of 4062F
AC.
To perform the 4062F AC Test, pin boards must be installed in the SWM at
Note
the following slots at least:
slots: 4, 8, 12, 16, 20, 24, 28, 32, 36, 40, 44, 48.

Required Equipment

Pulse test module (HP 16354F)
Digitizing oscilloscope (HP 54121T)
HP-IB cable
20 dB attenuator
SMA(m)-to-BNC(f) adapter
BNC(m)-to-BNC(m) cable (less than 1.5 m)
The HP 54121T oscilloscope must be warmed up for at least one hour.
Note
When you execute the PV4062F.AC Test, be sure that the HP 16354F is
Warning
removed from the HP 4085F SWM. Then mount the HP 16354F on the SWM
according to messages displayed by the system software.
If you mount the HP 16354F on the SWM when a voltage or current source
is forcing voltage or current, dangerous voltages may be present on the
connectors of HP 16354F, and may shock you.
6-40 System Verification