Command format: Fixed Length
Communication direction
From external device to
PV310
Response: Normal
Response: Error 1
Response: Error 2
•
ERROR signal = "ON"
•
Error Code: E0175 is
displayed on the screen
To specify the command format (Comma separated/Fixed Length), select "ENVIRONMENT" > "Serial
Communication" > "Command" from the menu bar.
Command / Description
%
T
?
?
a
a
??
01 to 08
aaaa
0001 to 0254
or
0001 to 0254
bbbb
0001 to 0255
or
0001 to 0255
0000 - 0255: Unused digits are filled with '0'.
0000 - 0255: Unused digits are filled with 'Blanks'.
%
T
?
?
a
a
Send the same command as the sending command to PV310 from external device.
%
U CR
Cause:
Command in the unspecified form or not provided was received.
Examples:
-
%T1,80,255CR (any other checker number than 01 - 08 was specified)
-
%B01CR (not provided command)
%
Z CR
Causes:
•
The max. value of Slice Level specified by a sending command is smaller than the
min. value of Slice Level.
•
The max. value of Slice Level is equal to the min. value of Slice Level.
•
Max. /min. values of Slice Level specified by sending command are not within the
range between 0000 and 0255.
a
a
b
b
b
b CR
Specify the number of Slice Level in 2-digit form.
Slice Level
A
B
Specified number 01 02 03 04 05 06 07 08
Specify the number of Slice Level-Min. in 4-digit form.
Specify the number of Slice Level-Max. in 4-digit form.
a
a
b
b
b
b CR
C
D
E
F
G
H
269