6.18 Rotation Adjustment (Gray Edge Horiz., Gray Edge
Vert.)
6.18.1 Setting Procedure
The detection and setting procedures of Area 1 and Area 2 are the same for Gray Edge. (See page 86.)
In this section, the differences between Rotation Adjustment (Gray Edge Horiz., Gray Edge Vert.) and
Gray Edge and the way of testing are described below.
Detection Position
Position Adjustment (Gray Edge) cannot set "Detection Position". Always detects the nearest edge from
the start point.
Testing Areas 1 and 2
Before registering the base position, you can perform a test to check if the Areas 1 and 2 detect an edge.
When the live image is displayed, perform the test after importing a new image. When a memory image is
displayed, perform the test using the memory image currently displayed without import of a new image.
1.
Select "ROT. ADJ. (GRAY EDGE)".
2.
Press the A key while the tab or the item is highlighted.
If no edge was detected, the following error number is displayed in the message area.
E0044
Check the image to make sure that the error code above does not appear or the edges are detected
Try to set to the same width between the Area 1 and Area 2. It is advisable to minimize the
width value as much as possible.
You cannot assign different memories to the horizontal and vertical areas.
Wrong example:
Horizontal area=memory 1, Vertical area=memory 2
You can assign different groups to the horizontal and vertical areas.
Correct example:
Horizontal area=group A, Vertical area=group B