Procedure
To back away from an ultimate limit:
1. Set the Ultimate Stop Override switch on
the safety board ON (downwards).
The red Ultimate Stop Disabled LED is lit.
2. Move the axis away from the ultimate stop
until the position is within the software
limits and the corresponding green LED
on the safety board is lit.
3. Press the Ultimate Stop Reset button on
the safety board to reset the ultimate stop
relay.
The red Ultimate Stop Relay LED must go off,
otherwise you are not free of the ultimate
limit.
4. Deactivate the Ultimate Stop Override
switch.
The red Ultimate Stop Disabled LED goes off.
If the system is not positioned within the soft-
ware limits, you can only operate the system if
SW8 on both motion controllers is active (up-
ward position).
Note. After verifying ultimate limits, it is im-
portant to reset both motion controllers and
restart the CardioMD application software.
5. Reset both motion controllers (SW10 in
position up).
6. Then remove the reset on the master mo-
tion controller (motion controller to the
right – SW10 in position down).
7. Finally, remove the reset on the slave mo-
tion controller (motion controller to the
left – SW10 in position down).
If the CardioMD application is running on the
acquisition PC, close down the application and
then restart it:
8. Ensure that the acquisition PC is display-
ing the Persistence page. Click the Close
1MAN0158-H11 04-2005 CSIP Level 1
Details
Ultimate Stop
Reset button
Ultimate Stop
Relay LED
Ultimate Stop
Disabled LED
Ultimate Stop
Override switch
Vertical Ultimate
Stop OK LED
Rotate Ultimate
Stop OK LED
Horizontal Ultimate
Stop OK LED
CardioMD Service Manual 4-47
Calibration