-event {value}
Time
ERRLOG
1MAN0158-H11 04-2005 CSIP Level 1
Example:
-> DYNCOR
Dyncor: 0:HiDyn = 300, 1:LoDyn = 160, 2:Thresh =
400
Enable/disable event transmission to the PC. The following values apply:
0: Disable event transmission.
1: Send uncorrected events.
2: Send energy corrected events.
3: Send linearity corrected events.
4: Send energy and linearity corrected events.
On power-up and reset, the correction tables are initialized with zeros,
which is equivalent to 'uncorrected'.
This is a useful utility when troubleshooting detector performance
problems. This utility allows for specific corrections to be loaded when
acquiring an acquisition to determine if a failure is in the correction table or
detector.
Example:
Event 1
Note. Remember to execute this command after the tables have been
loaded when accessing an acquisition screen.
Identifies the time from when the EDC board (system) was last reset.
Example:
-> TIME
System time is 404658 seconds since reset
Using the logging facility it is possible to get information with various
degrees of details. Default setting is all logging off.
Example:
< errlog 0x1000 > = CanBus communication
< errlog 0x2000 > = CanBus HW errors
< errlog 0x4000 > =
The logging setting is the result of an OR operation performed on these
individual hexadecimal values. For example, spcrifying
enabes logging of CanBus communication, CanBus HW errors and
FireWire input.
FireWire input
errlog 0x7000
CardioMD Service Manual 5-39
Diagnostics