Final System Calibration And Performance Testing; Checking Status Led's - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

8.4 Final System Calibration and Performance Testing

Refer to Chapter 3 Getting Started in the CardioMD Operator's Manual for information on system
startup.
Once the CardioMD system has been installed and powered up, the laptop PC has been powered
up and the CardioMD application software is running, the following checks must be performed in
order to verify the correct operation of the CardioMD system.
8.4.1

Checking Status LED's

Check that the three LED's on the LED panel next to the acquisition PC correctly indicate the
state of the CardioMD system:
Procedure
After the CardioMD has been powered up:
1. Ensure that Collision Override has been
pressed as part of normal power-up
procedure.
2. Check that the Collision LED and the E-
Stop LED are both OFF.
3. Check that the green Power OK LED is
ON.
4. Activate one of the system's collision
sensors.
5. Check that the yellow Collision LED is
ON.
6. Check that the green Power OK LED is
OFF.
7. Press Collision Override on the hand
controller.
8. Check that the yellow Collision LED is
OFF.
9. Check that the green Power OK LED is
ON.
10. Activate one of the E-Stop buttons.
11. Check that the yellow E-Stop LED is ON.
12. Check that the green Power OK LED is
OFF.
13. Reset the E-Stop button and press
Collision Override on the hand controller.
14. Check that the yellow E-Stop LED is
OFF.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Installation
CardioMD Service Manual 8-87

Hide quick links:

Advertisement

Table of Contents
loading