Procedure
5. Slide in the base cover from the front.
Fasten the cover:
6. Mount 4 screws (and washers) from the
left-hand side into the 4 stays.
7. Mount 2 screws (with washers) from the
right-hand side into the base frame.
8. Re-mount the cover on the gantry tower.
9. Re-mount the tower top cover.
10. Mount the rails for the sliding cover onto
the base, using 4 pcs. M6 screws and a
5 mm Allen wrench.
11. Connect the sliding cover to the tower
using a 2 mm Allen wrench.
12. Mount the stay for the cable tray in the
M12 hole to the left on the rear side of the
gantry base.
1MAN0158-H11 04-2005 CSIP Level 1