31. Protect cables on the floor.
32. Perform local area network and DICOM setup.
33. Configure regional settings, institution name and units for patient height and weight.
34. Set up hardcopy printer.
35. Verify proper operation and detector performance.
36. Recalibrate detector if found required when checking performance (see Chapter 4
Calibration).
37. Acquire customer requested uniformity correction tables, etc.
8.3.2
Tools
8.3.2.1
Tools Shipped with System
•
2 dolly wheel bars
•
2 jack beams
•
4 plastic protection feet
•
4 socket head cap screws M12 × 75 mm
•
1 rectangular plywood plate (plate for trolley) 24 × 24 inch (60 cm × 60 cm)
•
1 ½" ratchet
•
1 leveling tool (special screw driver for the leveling screws)
•
2 M10 × 200 mm long bolts
•
1pcs 15 mm masonry drill.
8.3.2.2
Tools not Shipped with System
•
Pallet Jack
•
Spirit level with tolerance +/– 0.05°
•
Hammer drill
•
Metric Allen keys, 1.5 through 10 mm
•
3/8" drill with Phillips heads bit (used for disassembling shipping containers)
•
Phillips screwdriver
•
3 mm standard screwdriver
•
A set of Torx keys in the range of T6 – T25
•
Wrenches: 16 mm, 17 mm and 19 mm
•
Mandrel used to set the anchors.
1MAN0158-H11 04-2005 CSIP Level 1
Installation
CardioMD Service Manual 8-23