Procedure
6. Take out one transmission scanner, and
then the other, from inside the box.
You are now ready to start the installation of
the CardioMD AC option as described in the
following sections.

12.6.2 Electrical Installation

12.6.2.1 Modifying Detector Cabling

Procedure
1. Follow the procedure provided in
Chapter 3 Getting Started of the
CardioMD Operator's Manual to shut
down the system.
2. When the system is powered down,
disconnect the mains cable from the mains
inlet on the rear of the table console.
3. Remove the detector top cover.
4. Disconnect the input power connector
from the detector power supply.
1MAN0158-H11 04-2005 CSIP Level 1
Details
WARNING
Be careful when handling the transmission scanners – they are quite heavy. Do
not drop a transmission scanner. Dropping a transmission scanner may cause
personal injury or damage the equipment.
Details
CardioMD AC Option
CardioMD Service Manual 12-91