12.6.3 Mechanical Installation

12.6.3.1 Preparing Detector Covers
Procedure
1. Remove the pre-cut knock-out pieces from
both ends of the detector top cover, using
an appropriate pair of pliers.
Caution. Be careful when removing the
knock-out pieces. Using excessive force may
damage other parts of the cover.
2. Dismount the two detector end covers.
3. If necessary, use a small hammer to
remove the three pre-cut knock-out pieces
in both covers.
4. Take care not to damage the detector
cover outside of the knock-out pieces.