Philips CARDIOMD Service Manual page 418

Table of Contents

Advertisement

Repair Procedures
Procedure
7. If required, turn the motor shaft by hand
(turn the wheel located next to the motor)
to translate the line souce housing to a
more convenient position inside the
compartment.
8. Use a 2.5 mm Allen key to loosen and
remove the 8 screws holding the line
source housing to the two slide bearings.
9. Push the slide bearings out from under the
line source housing.
10. Carefully turn the line source housing in
the direction indicated on the photo and
lift it out of the compartment.
Note. To prevent damage to housing shutter
components, it is important to turn the line
source housing in the indicated direction when
removing the source housing.
12-70 CardioMD Service Manual
Details
1MAN0158-H11 04-2005 CSIP Level 1