Procedure
reference scan, wait until the the line source is
at center position and then stop the reference
scan. Proceed as follows:
2. On the Persistence page, click the Tools
button to display the Calibration &
configuration menu.
3. Click the Daily reference scan button.
The Daily Reference Scan page appears.
4. Set Termination counts to 1000000 (1
million).
5. Click Start to start the reference scan.
6. Wait until the line source of the
transmission scanner that is to be aligned
is translated to center position inside the
line source compartment.
Note. This procedure is most convenient for
transmission scanner 2. When aligning
transmission scanner 1, you need to wait for
transmission scanner 2 to complete its entire
reference scan before transmission scanner 1
moves to center position.
7. Click Stop to stop the reference scan.
8. Click Exit and then OK to interrupt the
reference scan and close the Daily
Reference Scan page.
9. In the Calibration & Configuration menu,
click Exit to return to the Persistence page.
Alternatively, the line source housing can be
moved to center position by means of software
commands sent to the line source motion
controller via an RS 232 connection to the
EDC board. See the section Manual Control of
Line Source Motions, page 12-24 for details.
Next, set up a static acquisition:
10. On the Persistence page, click the Static
button.
1MAN0158-H11 04-2005 CSIP Level 1
Details
CardioMD AC Option
CardioMD Service Manual 12-83