Gantry Software Updates
Preparing for Firmware Upgrade
Procedure
1. Remove the detector top cover to access
the EDC modules. See instructions in
Chapter 6 Repair Procedures.
2. Remove the ECG cable from the ECG port
labeled A1 RS232.
3. Connect one end of the serial interface
cable to this port.
If you are using a laptop acquisition PC:
4. Use the USB to RS 232 adapter to connect
the opposite end of the serial interface ca-
ble to one of the acquisition PC's USB
ports.
See section 7.2.2, page 7-2 for instructions on
determining which COM port you are using.
If you are using a tower acquisition PC:
5. Connect the opposite end of the cable to
the acquisition PC's COM1 serial port.
6. Insert the software installation CD-ROM
in the acquisition PC's CD-ROM drive.
Upgrading Micro Controller Firmware
Procedure
To load micro controller firmware to the EDC
Board, you must start the Flash program. Pro-
ceed as follows:
1. In the Windows Explorer, select the folder
EDC CardioMD programming for
Module rev. 9CSY0900-B02 or
on the CD-ROM.
later
2. In this folder, double-click the icon
to start the Flash program.
flash.exe
7-8 CardioMD Service Manual
Details
III
I, II
I, II
III
Details
Serial interface
cable
ECG cable
1MAN0158-H11 04-2005 CSIP Level 1