Procedure
9. On the Persistence page, click the Tools
button.
10. When the Calibration & Configuration
menu appears, click Service.
11. If prompted, enter the service password
(currently DDD).
12. In the Service Calibrations menu, click
Node configuration.
13. In the Nodes Configuration dialog,
double-click the
to add a new node.
+
The Add a new node dialog appears.
14. Select the FOLDER option and click
Select.
15. Check off import/export capabilities from
the node as shown: Yes to Export and No
to Import.
Note. If you want to both import and export
from the node, check off Yes in both
selections.
16. To select the folder to where files will be
exported, click the ... button.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Diagnostics
CardioMD Service Manual 5-45

Hide quick links:

Advertisement

Table of Contents
loading