Mounting The Table Console To The Gantry Frame; Mounting Anchors For Table Console - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Installation
Procedure
17 mm wrench.

8.3.3.14 Mounting the Table Console to the Gantry Frame

Procedure
1. Using the long bolts as guides, push the
table base in position. A key in the gantry
frame will guide the base the last ¼"
(5 mm).
2. Remove the long screws and replace them
with M10 × 20 mounting screws, one at
the time.
3. Using an 8 mm Allen wrench, tighten the
screws and loosen each screw again by
approximately half a turn of the wrench.
4. Lower the rear end of the table base and
remove the jack beam.

8.3.3.15 Mounting Anchors for Table Console

Procedure
1. Using a 10 mm Allen key, mount 2 M12 ×
30 anchor screws with washers at the rear
end of the table. Tighten the bolts.
2. Also tighten the 2 anchor bolts at the front
end of the gantry.
8-36 CardioMD Service Manual
Details
Details
Details
1MAN0158-H11 04-2005 CSIP Level 1