Table 3.1 Continuous outputs
Connector Name
OC2
OC3
C5
C6
Table 3.2 Safety interrupted outputs
Connector C8 goes to the Safety Stop circuit. This input is used to interrupt power to the motors in
case of collision, motion controller software fault, or if one of the E-Stop buttons is pressed.
The electronics within CardioMD, including the detectors, are operated on uninterruptible 55 V
DC. Local DC-DC converters internal to the electronic modules convert the 55 V DC supply
voltage to the voltage levels used internally (i.e., 3.3 V DC, +/– 5 V DC, 12 V DC, +/– 15 V DC
and HV).
The AC power supply contains a number of fuses. The access and location of these fuses are
indicated in Figure 3.6 on page 3-6. The electrical location of the fuses is shown in Figure 3.7
below.
MAINS fuses
Figure 3.7 Electrical location of fuses in power supply
1MAN0158-H11 04-2005 CSIP Level 1
Voltage
Function
115V AC
Series III: ACQUISITION PC and ECG gate
Series I and II: ACQUISITION PC
55V DC
SPARE
55V DC
DETECTOR 1 EDC BOARD AND DETECTOR PSU
55V DC
DETECTOR 2 EDC BOARD
55V DC
BOTH MOTION CONTROLLERS
ROTATION AND VERTICAL AMPLIFIER*
24V DC
SAFETY BOARD
Voltage
Function
180V DC
ROTATE DRIVE AMPLIFIER
180V DC
Y DRIVE AMPLIFIER
55V DC
X DRIVE AMPLIFIER
24V DC
TABLE DRIVE
SECONDARY FUSES
ISOLATING TRANSFORMER
115V
to PC
180 V POWER
CORRECTION
55 V POWER
CORRECTION
CardioMD Service Manual 3-7
Detailed Description