Installation
Procedure
The ECG gate connects to the panel on the
rear of the table console.
3. Remove the two screws from the plate
covering the ECG gate power output
connector.
4. Remove the plate.
5. Use the IEC power cable and the BNC
cable supplied with the CardioMD system
to connect the ECG gate to the system.
6. Set the trigger polarity to negative.
The trigger polarity switch is located at the
rear panel of the gate.
8-72 CardioMD Service Manual
Details
ECG
trigger
ECG gate
power cable
1MAN0158-H11 04-2005 CSIP Level 1