Philips CARDIOMD Service Manual page 417

Table of Contents

Advertisement

Procedure
The photo on the right shows the line source
housing inside the compartment.
4. Locate the transportation screws sitting in
the spare holes in the line source housing.
5. Move the two transportation screws to the
center holes as shown in the photo on the
right.
6. Tighten the transportation screws so that
they lock the shutter in the closed position.
CAUTION
RADIOACTIVE
MATERIAL
1MAN0158-H11 04-2005 CSIP Level 1
Details
WARNING
Make sure that the transportation screws are in place and locking the shutter
before attempting to remove the source holder. Failing to lock the shutter may
cause you to be exposed to radiation from the source inside the holder.
CardioMD AC Option
Transportation screws
Transportation screws
CardioMD Service Manual 12-69