Philips CARDIOMD Service Manual page 421

Table of Contents

Advertisement

Procedure
4. Carefully push the slide bearings into
position underneath the line source
housing.
5. Use a 2.5 mm Allen key to mount the 8
M3 × 30 Allen screws, securing the line
source housing to the slide bearings.
Note. Before fastening the screws, check that
the line source housing is perpendicular to the
rails, e.g. by aligning one of its sides with
screw heads in the two rails.
If required, turn the motor shaft by hand to
translate the line souce housing to a more
convenient position inside the compartment.
The photo on the right shows the line source
housing installed on the slide bearings and
with all screws mounted.
6. Finally, remove the two transportation
screws securing the line source housing
shutter mechanism.
1MAN0158-H11 04-2005 CSIP Level 1
Details
CardioMD AC Option
Transportation screws
CardioMD Service Manual 12-73