Mains Voltage Configuration - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Procedure
19. To prevent cable connections from being
stressed, attach two large Ty-Wraps to the
cable on the inside of the cabinet.

8.3.3.23 Mains Voltage Configuration

Procedure
1. From the rear of the table console, remove
the rectangular plate on the AC power
supply.
2. Place the plug in the socket with the
closest match to the actually measured
mains voltage.
3. Verify that both fuses match the rating at
the selected socket. Leave the alternatively
rated fuses in the plastic bag in the bottom
of the voltage selector compartment.
If configuring power for the final installation,
the remaining fuses should be discarded. This
is to ensure they are not mistakenly used as a
replacement fuse as they will have the wrong
value for the power setting.
4. Remount the rectangular cover plate.
Note.
Nominal mains voltage
90 – 107 V AC
108 – 132 V AC
180 – 208 V AC
209 – 228 V AC
229 – 264 V AC
5. Connect the CardioMD system's power
cord to the mains supply.
Note. The power cord provided with the system is an unterminated cord. The mains wiring is
approved to be supplied, either via a plug into a wall outlet or hardwired to the installation.
Procedure
6. Power up the system.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Details
Setting
100 V
120 V
200 V
220 V
240V
Details
100V
120V
200V
220V
240V
12A
12A
6.25A
6.25A
6.25A
FUSE
I
0
FUSE
CardioMD Service Manual 8-65
Installation
Mains voltage
setting
SETTING
FUSE RATING
( both fuses)