Procedure
14. Take the two fixing screws out of the
FireWire connector and hold it against the
internal connector panel from the rear of
the system.
15. Mount the fixing screws from the front
side of the system.
16. From the front side, insert the FireWire
cable from the FireWire hub into the
mounted FireWire connector.
17. Plug the 2-way M-N-L connector marked
D12 into connector D12 on the safety
board.
18. Attach the BNC connector to the BNC
adapter (for the ECG gate) on the rear
connector panel.
Note. You may have to remove the connector
panel to access the BNC adapter.
Three yellow cables are equipped with DB9
connectors with code plates mounted. Each
code plate will only fit in one position over the
three DB9 connectors on the safety board.
19. Connect the three cables at the correct
positions and fasten the socket-headed
retaining screws.
20. Plug the three cables with 3-way M-N-L
connector into the mating connectors on
the safety board.
"R" goes to D15, "H" goes to D14 and "V"
goes to D13.
1MAN0158-H11 04-2005 CSIP Level 1
Details
Mount fixing
screws here
Installation
Connector D12
CardioMD Service Manual 8-41