Installation
Procedure
12. When the DB15 connector is in place,
insert the collision circuit jumper
(9PLG1345) supplied with the AC option.
Note. Installation of the collision circuit
jumper is required in order to allow the
CardioMD system to move before the
transmission scanners are installed.
13. On both EDC boards, remove the cover
collision board (9COL0899) and replace it
with the EDC AC IF board (9ACI1312).
14. Reestablish the cable connections existing
previously to the new board.
15. Plug the 6-way plug from the AC Option's
detector interface cable (9CBL1313) into
the mating header.