Mounting Transmission Scanners; Aligning Transmission Scanners - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Procedure
4. Again, press down each collision sensor at
the other end of the line source
compartment, one by one, easing the cover
into place.
5. Remount the edge cover and fasten it
using the two screws previously removed
from the service cover.
6. Repeat the procedures outlined above for
the other transmission scanner.

12.5.3.11 Mounting Transmission Scanners

Instructions for mounting the transmission scanners on the detectors are provided in section
12.6.3.4 on page 12-103.

12.5.3.12 Aligning Transmission Scanners

The AC option transmission scanners must be aligned:
When the AC option is installed for the first time
Whenever a line source compartment has been detached from the AC option bracket
Whenever a line source housing has been removed from the line source compartment
After installation of new line sources
When the CardioMD system issues an error message saying "Line source offset has exceeded
+/– X mm".
Note. Aligning transmission scanners is performed with the transmission scanners mounted on the
detectors.
1MAN0158-H11 04-2005 CSIP Level 1
Details
CardioMD AC Option
CardioMD Service Manual 12-81