5. In order to protect the 3 cables, make sure
they are safely placed on the gantry base.
6. Pull out the pallet.
7. Using an 8 mm Allen wrench, loosen the 2
M10 screws in each end bar and remove
the mounting plates in each end.
8. Leave the plate in the middle.
9. Raise the gantry to app 8" (20 cm) above
the floor level and mount a wheel bar in
each end by using the same screws in the
same holes used for the mounting plates.
10. Lower the jack until the gantry is standing
on the four wheels.
11. If required, remove the jack beams.
Note. The Gantry on its Dolly with the jack beams mounted can go through 91 cm (3 ft.) door-
ways.
Without the jack beams, the gantry can pass a doorway of 75 cm (2.5 ft.).
8.3.3.4

Determining the Actual Position for the System

Note. Use the template shipped with the system to determine the actual position of the gantry.
The template shipped with the system (see Figure 8.17 page 8-30) provides information on the
minimum distance to the walls, the minimum room layout for installation with a small table
(supine patients only) and a large table (supine and prone patients), as well as the preferred