Diagnostic Software Utilities
5.6.3

FireViewer

This program is part of the FireWire driver installation on the acquisition PC. The program
enables you to perform a quick check of the integrity of the FireWire communication.
Procedure
To access the program:
1. Click the Windows Start button, and then
select Program, ubCore, FireViewer.
The FireViewer program is located in the
directory
c:\Program
Files\Unibrain\ubCore
When the FireViewer program has been
started, it immediately shows the status of the
communication.
In CardioMD Series III systems, the FireWire
comprises four nodes:
The controlling node, i.e. the acquisition
PC, shown as a white box
The FireWire hub shown as a hatched box
(normal operation)
The two regular nodes, i.e. the two
detector EDC modules, shown as green
boxes.
The FireViewer window shown in the example
indicates normal communication.
5-40 CardioMD Service Manual