Repair Procedures
Procedure
20. Move the handle to the unlocked position
to mechanically detach the transmission
scanner from the detector.
Note. The label on the AC Option bracket
indicates the locked and unlocked position of
the AC Option handle.
21. Lift the transmission scanner free of the
detector end.
Procedure
22. Locate the collision circuit jumper (part
no. 9PLG1345) and install it on the
connector behind the port on the detector
cover.
Note. The collision circuit jumper is only
needed if the CardioMD system is to be used
without the AC option. If the transmission
scanners are just being removed for service
purposes, you do not need to install the
collision circuit jumper.
23. Repeat steps 19 to 22 to dismount the
transmission scanner on the other detector.
12-32 CardioMD Service Manual
Details
WARNING
Be careful when dismounting the transmission scanner from the detector. The
transmission scanner weighs more than 40 lbs.
Details
WARNING
Make sure to handle the transmission scanners carefully. Do not drop a
transmission scanner. Dropping a transmission scanner may cause injury or
Handle in unlocked position
1MAN0158-H11 04-2005 CSIP Level 1