Gantry Software Updates
Procedure
10. Click Select Hex File and browse the
folder on the CD-ROM to find the file to
download.
The file name is of the form
, where
7MOT0096Xnn.h86
of the software.
11. Click Open.
12. Select the COM port used.
13. Click Program Target Device
The program issues the message:
Please ensure that the Target
Board is in Bootstrap Loader Mode
When you see this message:
14. Set SW10 on the motion controller in po-
sition UP to reset the motion controller.
15. Set SW9 in position UP to set the motion
controller in program mode.
16. Set SW10 back to DOWN position to
remove the reset.
17. Click OK to download the software to the
motion controller.
A progress bar should now appear.
Note. If an error message appears instead of
the progress bar, press OK and repeat from
step 13 to try again. Refer to the
file in the
folder on the CD-ROM for fur-
MC
ther details.
18. Wait until the progress bar disappears and
a completion message appears. Click OK.
19. Set SW9 on the motion controller back in
the DOWN position.
20. Set SW10 UP and then DOWN to reset
the motion controller.
To update other motion controller:
21. Connect the SW load cable to the other
motion controller.
22. Repeat step 13 to 20 to download software
to the other controller.
23. Click Exit to close the Flash program.
7-16 CardioMD Service Manual