12.3.2.2 Gantry

When the CardioMD AC option is installed, a dedicated power supply is installed inside the table
console. This entails some modifications in the gantry cabling as illustrated in Figure 12.7.
115 V AC TO
CONNECTOR PANEL
24 V ES TO TABLE
BOARD
9CBL0887-
Figure 12.7 Modifications to gantry cabling when installing the AC option power supply
The continuous and safety interrupted outputs from the system power supply are now as shown in
Table 12.1 and Table 12.2 below:
Connector Name
OC1
C1
C2
C3
C4
C7
* if these are of the type that requires 55 V DC input
Table 12.1 Continuous outputs
Connector Name
OC2
OC3
C5
C6
Table 12.2 Safety interrupted outputs
12-10 CardioMD Service Manual
EXISTING CABLES
3
2
1
1
2
3
Voltage
Function
115V AC
ACQUISITION PC AND AC OPTION POWER SUPPLY
55V DC
SPARE
55V DC
DETECTOR 1 EDC BOARD AND DETECTOR PSU
55V DC
DETECTOR 2 EDC BOARD
55V DC
BOTH MOTION CONTROLLERS
ROTATION AND VERTICAL AMPLIFIER*
24V DC
SAFETY BOARD
Voltage
Function
180V DC
ROTATE DRIVE AMPLIFIER
180V DC
Y DRIVE AMPLIFIER
55V DC
X DRIVE AMPLIFIER
24V DC
TABLE DRIVE AND AC OPTION POWER SUPPLY
48 V to AC options
AC-DC PSU
XP SDS120PS48B
115 V AC 1
115 V AC 2
+24 V ES
24 V RET
NEW CHASSIS
(UPDATE KIT)
CardioMD AC Option power supply
Existing connector
inside Table console
NEW CABLES IN CONSOLE
(UPDATE KIT)
1
2
3
40 CM
40 CM
3
2
1
24 V E-ST
MAIN PSU
1MAN0158-H11 04-2005 CSIP Level 1
PSU
OC1
3
2
1
C6
1
2
3